Whilst all efforts are made to provide accurate and timely information from the relevant source/documentation, please be aware that the information supplied may not be the most current version. The accuracy of the detail has not been confirmed by the Department and therefore should not be relied upon without first confirming the contents.

Verify a liquid measuring instrument using volume measures

Verify a liquid measuring instrument using volume measures

Unit of competency
National Code
MSATMVER402A
State Code
S7284
TGA Status
Replaced
DTWD Status
Replaced
Current Release Number
1.00
Current Release Date
07/04/2011
State Implementation and Classification
Approved Date
19/08/2014
Field of Education
030305 - Materials Engineering
Original Release Date
19/08/2014
Nominal Hours
60
Description
This unit of competency covers the ability to apply National Test Procedures to determine whether a liquid measuring instrument using volume measures is suitable for trade and then mark it accordingly. It also involves the installation and/or repair of liquid measuring instruments.
Notes
Elements and Performance Criteria
1. Install and repair liquid measuring instruments
  • 1.1. Access and correctly interpret information for installing or repairing instruments
  • 1.2. Select required components, tools and equipment in accordance with manufacturer/component supplier specifications
  • 1.3. Perform installation or repair in accordance with legislation, industry codes of practice and organisational guidelines
  • 1.4. Avoid instrument damage during installation or repair
2. Prepare for verification
  • 2.1. Identify and evaluate the type of instrument to be verified
  • 2.2. Access and correctly interpret documentation required for the verification
  • 2.3. Identify and access test equipment, products and consumables required for the verification
  • 2.4. Liaise with the trader to provide the items required on site
  • 2.5. Ensure specified test equipment is fit for purpose in accordance with applicable legislation and organisational procedures
  • 2.6. Store and transport test equipment in accordance with organisational procedures and industry best practice
  • 2.7. Identify relevant local workplace health and safety issues and implement appropriate control strategies
3. Evaluate liquid measuring instrument performance
  • 3.1. Evaluate whether the operating environment will impact on the instrument performance
  • 3.2. Modify the operating environment or implement alternative arrangements to ensure reliable test conditions as necessary
  • 3.3. Identify the maximum permissible errors for the instrument from the legislative requirements
  • 3.4. Use test equipment safely in accordance with applicable legislation and organisational procedures
  • 3.5. Check instrument for compliance with the appropriate Certificates of Approval
  • 3.6. Inspect the instrument in accordance with relevant National Test Procedure and appropriate National Measurement Institute policy
  • 3.7. Evaluate results against prescribed performance criteria and determine if the instrument is suitable for trade use in accordance with legislative requirements
4. Report verification results
  • 4.1. Display the verification result on the instrument in accordance with legislative requirements
  • 4.2. Use test reports to present verification results in the required format
  • 4.3. Complete verification documentation in accordance with legislative requirements and organisational procedures
  • 4.4. Communicate results within the specified time and in accordance with organisational guidelines
5. Model and encourage compliance with statutory requirements
  • 5.1. Apply organisation's procedures and practices to meet licensing authority's requirements
  • 5.2. Clarify any issues about licensing requirements with the licensing authority as they arise
  • 5.3. Review work and seek feedback from others to confirm continuing compliance with licensing requirements
  • 5.4. Identify implications of non-compliance with licensing requirements
  • 5.5. Identify inadequacies in trader's procedures and practices which may contribute to non-compliance with licensing requirements and/or national measurement legislation
  • 5.6. Explain inadequacies and possible remedial actions to trader
6. Maintain statutory records
  • 6.1. Keep accurate and complete records in accordance with licensing requirements
  • 6.2. Ensure authorisation, training and relevant licences are current in accordance with organisational and licensing requirements
  • 6.3. Inform the licensing administering authority of changes to personal information as required by statute
RANGE STATEMENT
The range statement relates to the unit of competency as a whole. It allows for different work environments and situations that may affect performance. Bold italicised wording, if used in the performance criteria, is detailed below. Essential operating conditions that may be present with training and assessment (depending on the work situation, needs of the candidate, accessibility of the item, and local industry and regional contexts) may also be included.
Prescribed performance criteria for instruments
Prescribed performance criteria for instruments may include:
design is in accordance with the appropriate Certificates of Approval
performance meets the criteria described in the Certificates of Approval, National Test Procedure and legislation
Appropriate documentation
Where reference is made to documentation, it is expected the latest version will be used.
Appropriate documentation may include:
reference standards
Certificates of Verification
Certificates of Approval for liquid measuring instruments
test procedures for verifying liquid measuring instruments
organisational test reports
organisational procedures e.g. company quality assurance manual
National Measurement Act
occupational health and safety (OHS) regulations, guidelines and procedures and material safety data sheets (MSDS)
equipment manuals and warranty, supplier catalogues and handbooks
Certificates of Approval
Certificates of Approval may include:
any Certificate issued by the Chief Metrologist under Regulation 60 of the National Measurement Regulations approving the pattern of a liquid measuring instrument as being suitable for trade
Test equipment
Test equipment may include:
reference standards of measurement
equipment other than reference standards of measurement such as pumps, funnels and hoses
Legislation
Legislation may include:
national measurement legislation
applicable Commonwealth, state and territory OHS legislation
National Measurement Institute policy
National Measurement Institute policy may include:
accepted test procedure variations
bulletin
instruction
determination
National Test Procedures for liquid measuring instruments
National Test Procedures for liquid measuring instruments may include:
fuel dispensers other than LPG dispensers
bulk flowmetering systems for liquid hydrocarbons other than LPG tested using a volume measure
milk flowmeters tested using a volume measure
any other test procedure prescribed by the National Measurement Institute
OHS and environmental management requirements
OHS and environmental management requirements refer to:
all operations must comply with enterprise OHS and environmental management requirements, which may be imposed through state/territory or Federal legislation. These requirements must not be compromised at any time
all operations assume the potentially hazardous nature of samples and require standard precautions to be applied
where relevant, users should access and apply current industry understanding of infection control issued by the National Health and Medical Research Council and State and Territory Departments of Health
Operating environmental impacts
Operating environmental impacts may include:
vibration
wind
heat
dust
electromagnetic interference
out of level
Records
Records may include:
test reports
safety procedures
a history of equipment calibration and test results
Liquid measuring instruments may include:
Verifiers may be licensed to test and mark specific classes of instruments from the following list.
Liquid measuring instruments may include:
fuel dispensers other than LPG dispensers
bulk flowmetering systems for liquid hydrocarbons other than LPG tested using a volume measure
milk flowmeters tested using a volume measure
any other liquid measuring instrument prescribed by the National Measurement Institute
EVIDENCE GUIDE
The Evidence Guide provides advice on assessment and must be read in conjunction with the performance criteria, required skills and knowledge, range statement and the Assessment Guidelines for the Training Package.

Overview of assessment
Competency must be demonstrated in the ability to perform consistently at the required standard for one class of liquid measuring instrument listed in the Range Statement.

Critical aspects for assessment and evidence required to demonstrate competency in this unit
Assessors should ensure that candidates can:
identify, access and apply test procedures
identify and use suitable reference standards
install and repair liquid measuring instruments to meet statutory requirements
evaluate and adjust the impact of the operating environment on the performance of the instrument
analyse test results to determine the instrument's suitability for verification (trade use)
maintain the security and confidentiality of data in accordance with organisational and regulatory requirements
report results in the required formats and expected timeframe.

Context of and specific resources for assessment
This unit of competency is to be assessed in the workplace or simulated workplace environment.
This unit of competency may be assessed with:
MSATMREF301AUse and maintain reference standards.
Resources may include:
liquid measuring instruments using volume measures, test equipment and reference standards
computer and relevant software and/or organisation information management system
Certificates of Approval for instruments
relevant legislative and organisational procedures.

Method of assessment
The following assessment methods are suggested:
questions to assess understanding of relevant procedures and remedial actions
review of verification reports prepared by the candidate
feedback from supervisors and peers regarding the candidate's ability to verify the performance of instruments in accordance with legislative and organisational procedures.
In all cases, practical assessment should be supported by questions to assess underpinning knowledge and those aspects of competency which are difficult to assess directly. Questioning techniques should suit the language and literacy levels of the candidate.

This competency in practice
Industry representatives have provided the case studies below to illustrate the practical application of this unit of competency and to show its relevance in a workplace setting.

A service station operator of a large company owned site requests a servicing company to visit their site as he has noticed that one fuel dispenser has a faulty display and the service station appears to have inconsistent dip reading. The verifier arrives and introduces himself and informs the owner that under their contract all fuel dispensers are due for checking. The verifier completes a safety check, AIP clearance form and asks about the location of the fuel return points. The verifier then blocks the lane next to the fuel dispenser, puts safety signs, barricades and test measures in place and ensures that the work area is safe.The verifier repairs the faulty display and proceeds to test the fuel dispenser. He visually checks the fuel dispenser against the Certificate of Approval and checks for any leaks noting any discrepancies on a verification report form. He then tests the fuel dispenser as per the National Test Procedures noting results on the test report and returning dispensed fuel to the storage tanks in a safe manner. The verifier finds the fuel dispenser is outside the maximum permissible error and adjusts the dispenser. After repeating the National Test Procedure, he finds the dispenser is within the maximum permissible error and then seals and marks the dispenser with his mark.
The verifier then tests the other fuel dispensers on-site. Some require adjustment before being verified and marked. Another dispenser is within the maximum permissible error and does not require adjustment. One dispenser cannot be adjusted within the maximum permissible error and the verifier removes the mark. After all dispensers are tested, the verifier completes the verification forms, fuel usage form and AIP form. The verifier informs the site manager that one dispenser cannot be used as it is non-compliant and issues a non-compliancy form. The site manager locks the dispenser until repairs can be made. The remaining forms are given to the site manager and the verifier departs.
Replaced By
State Code National Code Title Type
AXN89 MSMTMVER402 Verify liquid measuring instruments using volume measures Unit of competency